Home > Tendering > Production materials >

Hunan Chunguang Jiuhui Modern Traditional Chinese Medicine Co., Ltd. Stability Test Chamber Bidding Announcement

Release time:2023-10-30 00:00

Source:

According to the relevant provisions of the "Management Measures for Tendering and Bidding", our company will adopt an open bidding method for the procurement project of 4 stability test chambers. Qualified suppliers are welcome to participate in the bidding.

1、 Procurement project information

Buyer: Hunan Chunguang Jiuhui Modern Traditional Chinese Medicine Co., Ltd

Procurement Project Name: Stability Test Chamber (4 units)

2、 Description of equipment to be purchased

1. Please refer to the document for detailed equipment parameters. Additionally, two 15KW variable speed and voltage regulators are required.

2. The equipment has a warranty period of 2 years, and within 3 years after the warranty period, only accessory fees will be charged. (From the date of acceptance).

3. During the warranty period, after a malfunction occurs, the supplier is required to respond by phone within 30 minutes of receiving the report and arrange for technical personnel to arrive at the site within 36 hours to troubleshoot.

4. Payment terms: 30% advance payment will be made after the contract is signed. 30% payment will be made upon receipt of the full 13% value-added tax special invoice upon successful installation and commissioning of the goods. 30% payment will be made after 2 months of instrument operation and acceptance. 10% of the deposit will be used as a quality guarantee deposit. The full payment will be made in full after 12 months from the date of acceptance of the goods at the buyer's factory.

3、 Qualification Requirements for Bidders

1. Having independent legal personality, possessing the qualification level and registered capital required for the bidding project.

2. Capable of fully fulfilling the contract.

3. In the past three years, we have independently completed similar bidding projects without any quality or safety accidents caused by management reasons or administrative penalties.

4. Lawsuits without significant defendants, respondents, or executors may affect the performance of the bidding project contract.

5. Comply with other requirements set forth in national laws, regulations, and bidding documents.

4、 Composition of bidding documents

1. Quotation section (including detailed equipment configuration list, transportation, taxes, after-sales service, and payment methods).

2. Bring a set of copies (to be retained) of the following documents stamped with the official seal: copies of the business license and tax registration certificate of the enterprise; Original power of attorney from the legal representative, copies of the legal representative's ID card and the authorized person's ID card; After sales service commitment.

5、 Binding and submission of bidding documents

1. Each bidding unit shall prepare their own bidding documents according to the requirements of the announcement for bidding.

2. The bidding party shall bind and seal the quotation and information of the bidding documents in a booklet.

3. After stamping and sealing, please send it to our company by express delivery.

4. Bidding documents that exceed the deadline and have incomplete information are considered invalid.

5. On the day of bid opening, bidders must send personnel to the site to participate in the bid opening, otherwise it will be considered as a waiver.

6. Mailing address: Mr. Li, Procurement Department, No.10 Kangping Road, Liuyang Economic Development Zone, Changsha City, Hunan Province

6、 Bid evaluation and selection method: Adopt competitive negotiation and comprehensively select the best.

7、 Bid deadline: 9:30 am on November 7, 2023 (Beijing time)

8、 Bid opening date: 9:30 am on November 7, 2023 (Beijing time)

Bid opening location: No. 10 Kangping Road, Liuyang Economic Development Zone, Changsha City, Hunan Province

9、 Contact person for this project: Mr. Li, contact information: 15873338803

Technical parameter requirements for stability test box

Volume

1000L

Inner Dimension

900*800*1400

Overall Dimension

1080*1500*1750

Temperature Range

10-65℃

Temperature Fluctuation

±0.5℃

Humidity Range

20-95%

Humidity Deviation

≤±3%RH

Humidification Method

Fully Automatic

Operating Environment Temperature

5~40℃

Inner Tank Material

Mirror Finish 304 Stainless Steel

Outer Shell Material

Cold-Rolled Steel Plate with Powder Coating

Safety Devices

Power Overload Protection, Independent Working Chamber Over-Temperature Protection, Compressor Overload Protection, Over-Pressure Protection, Water Shortage Protection

Operating Power Supply

AC220V±10% 50HZ

Display Screen

Intelligent Touch Screen

Alarm System

Audible and Visual Alarm for On-Site Temperature and Humidity Deviation, Remote Alarm for Abnormal Conditions

Authority Control

Level 2 or Above (Login and Access), Audit Trail

Lighting Lamp

Equipped with Illumination

Ultraviolet Lamp

With Ventilation and Ultraviolet Disinfection Functions.

Printer

Equipped with a Mini Printer for Real-Time Printing of Temperature and Humidity Curves and Data

Data Storage

USB Storage Compatible

Loading Plates

5 Layers

3Q Certification

Complies with FDA, ICH and National Pharmacopoeia Requirements, and Provides Valid Calibration Report.

Special Requirements:

1. Conduct Tests at 2 Temperature and Humidity Points: 30-Minute No-Load Test, 24-Hour Load Test

2. Conduct Tests at 2 Temperature and Humidity Points: 30-Minute No-Load Test, 24-Hour Load Test (Approximately 3-4 Days)

3. Conduct Tests at N Temperature and Humidity Points: 30-Minute No-Load Test, 24-Hour Load Test, 3 Rounds of Challenging Tests (Door Opening/Closing, Power Interruption, etc.) (5-6 Days)

Energy Saving

Adopts Environmentally Friendly Fluorine-Free Refrigerant (R134a), Low Frequency, Low Energy Consumption, Energy Saving and Environmental Protection

After-Sales Service

2-Year Warranty with Free Maintenance During the Warranty Period. Response Within 24 Hours.

Uninterruptible power supply parameters (variable speed regulator)

Model: EK3C 15H

Input: 220/380Vac-240/415Vac, 50/60Hz

27.0A(Ph-N),3W+N+PE

Output: 220-240Vac, 50/60Hz, 1W+N+PE, 54.5A

Capacity: 15kVA/12kW

Battery voltage: 192VDC